Select "Force Calibration" in the pull-down menu under the "Calibration" menu in the menu of the electronic universal testing machine, and a dialog box will appear. At this time, enter the password to enter the calibration interface. In the first operation, the password is 123456, and you can directly select "OK". Then you will see the calibration interface on the right side of the interface. Press the [Calibration Settings...] button, and the test parameter modification interface will appear. Then select the force calibration settings, First enter the range, then select the number of calibration points, and then enter the calibration value you want. Press the "Start Calibration" button to enter the next interface, and you can start the calibration operation
Calibration method: Place the mouse on the point you want to calibrate in advance, and then you will see that the selected point to be calibrated has a black dotted box, which represents the selected point. When the point to be calibrated is approaching, press the left button of the mouse directly when calibration is needed. Then, when calibrating the next point, and so on. After all points are calibrated, the system will prompt that calibration is complete 摩擦磨損試驗(yàn)機(jī) 端淬試驗(yàn)機(jī) 拉力試驗(yàn)機(jī) 材料試驗(yàn)機(jī) 電子試驗(yàn)機(jī) 扭轉(zhuǎn)試驗(yàn)機(jī)
在電子試驗(yàn)機(jī)的菜單中選中"標(biāo)定"菜單下的下拉菜單中的"力值標(biāo)定",就會(huì)出現(xiàn)一個(gè)對(duì)話框,此時(shí)輸入密碼就可以進(jìn)入標(biāo)定界面,在第一次運(yùn)行時(shí)密碼:123456,直接選擇"確定"就可以了,然后你就會(huì)在界面的右邊看到的標(biāo)定界面,先按下[標(biāo)定設(shè)置…]按鈕,出現(xiàn)試驗(yàn)參數(shù)修改的界面,然后選擇力值標(biāo)定設(shè)置,先輸入量程,然后選擇標(biāo)定點(diǎn)數(shù),然后再輸入你要的標(biāo)定的值就可以了.按下"開(kāi)始標(biāo)定"按鈕后,就進(jìn)入下一個(gè)界面了,就可以進(jìn)行標(biāo)定操作了.
標(biāo)定方法:將鼠標(biāo)提前放在你要標(biāo)定的點(diǎn)上,此時(shí)會(huì)看到被選中的要標(biāo)定的點(diǎn)有一個(gè)黑色的虛線框,那個(gè)就代表選中的點(diǎn).在要標(biāo)定的點(diǎn)快要到的時(shí)候,需要標(biāo)定時(shí)直接按下鼠標(biāo)的左鍵.接著標(biāo)定下一個(gè)點(diǎn)時(shí),依此類推.所有的點(diǎn)都標(biāo)定完后系統(tǒng)會(huì)提示,標(biāo)定完成.